Determinação da acurácia do monitor oscilométrico Delta Life DL 1000 na mensuração da pressão arterial em cães anestesiados de diferentes faixas de peso

Correia, Bárbara SilvaMonteiro, Eduardo RaposoFerronatto, João Victor BarbieriQueiroga, Luciana BranquinhoBecerra, José Ricardo Herrera

Background: Arterial blood pressure is one of the most commonly variables monitored during anesthetic procedures in veterinary patients. The most reliable method for measuring arterial blood pressure in dogs and cats is the direct (invasive) method. However, the oscillometric method is less complex and more practical for clinical routine in small animals. Nevertheless, oscillometric monitors present great variability in accuracy. The present study aimed to determine the accuracy of the Delta Life DL 1000 oscillometric monitor for measurement of systolic, mean and diastolic blood pressures (SAP, MAP and DAP, respectively) in anesthetized dogs of different weight ranges. Materials, Methods & Results: This study was approved by the Institutional Ethics Committee of Animal Use. Fifteen female dogs of different breeds, weighing 11.6 ± 10.0 kg and with a mean age of 48 ± 51 months were used. All animals were scheduled for elective surgery under general anesthesia in the Institution Veterinary Hospital. Dogs were anesthetized with morphine, propofol and isoflurane and had a 20 or 22 gauge catheter introduced into the dorsal pedal artery for continuous, invasive monitoring of SAP, MAP and DAP. A blood pressure cuff was positioned over the middle third of the radius and connected to Delta Life DL 1000 monitor. Oscillometric readings of SAP, MAP and DAP were registered every 5 minutes, and invasive values were simultaneously recorded. Values obtained with both methods were compared (invasive versus oscillometric) by use of the Bland Altman method to determine the bias, standard deviation of bias and 95% limits of agreement. The percentages of errors between the methods within 10 mmHg and within...(AU)

Texto completo